Company Name :
Beijing Labonce Thermostatic Technology Co., Ltd.Brand :
LABONCEControlling Humidity Precision :
Humidity Fluctuation < ±2%RH;Humidity Deviation < ±3%RHControlling Temperature Precision :
Temperature Fluctuation < ±0. 5℃;Temperature Deviation < ±1.0℃Environment Temperature :
+5 ~ 35℃Power :
AC220V±10% 50HZAlarm system :
On-site beeping alarmCapacity :
800LHumidity :
20~95%RHReference Standard :
ICH2003、Chinese Pharmacopoeia 2020 editionLabonceqGS series largeqcapacity drug stability test chamber,with independent right and left door opening design.It issaasmall in depth and convenient for taking samples.The singleaside door opening minimizes the overall temperature and humidityachanges.It adopts highaquality craftsmanship design and selects imported highaquality components.The performance isastableaand reliable,suitable for users certified by GMP and CGMP.
1.Insulation material
Overall high-density polyurethane foam technology, with good insulation and moisture retention performance.
2.Chamber Materials
The exterior is coated with high quality steel plate,The liner is made of mirror stainless steel 304, no pollution source, easy to clean.
3.Control system
Programmable color touch screen controller;more than 7 inches,display frequency conversion output ratio.
4.Three Level Authority
Adopt the user name and password to log in. Different user names have different permissions at least three levels or above.
5.Audit tracking
The control system has historical alarm recording and operation recording functions, and supports exporting in an unchangeable file format using a USB flash drive.
6.Refrigeration system
Imported fully enclosed compressor;Save more than 50% of electricity and water.
7.Humidity control
Original imported capacitive humidity sensor, high precision, low drift, long service life, maintenance free.
8.Data management
Configuring needle micro printers and electronic data storage functions.
9.Safety device
Compressor overheat and overpressure overload protection, water shortage protection, dry burning protection system, independent overtemperature protection alarm system.
10.Double Door structure
Interior door tempered glass door,Open the outer door to observe the samples,The temperature and humidity inside the container will not change in a short time,The outer door is solid can keep temperature and humidity,It can also prevent the influence of external light.
11.Optional
Temperature- humidity deviation, sudden power failure , control by Remote SMS alarm.
Model
|
Temperature Range
( ℃ )
|
Humidity Range
(RH)
|
Capacity
(L)
|
Interior Dimensions(mm)
W×D×H
|
External Dimensions(mm)
W×D×H
|
Power
(kW)
|
Shelf
(Standard)
|
Remarks
|
Labonce-800GS-FC
|
10~65
|
20~95%
|
800
|
1200×490×1360
|
1385×895×1965
|
2.5
|
4/8
|
Customized
3000GS-FC
configuration
observation
window
|
Labonce-1000GS-FC
|
10~65
|
20~95%
|
1000
|
1400×510×1400
|
1615×925×1975
|
3.0
|
4/8
|
Our hours